These part numbers have the same form, fit, and function as pm1
Item Number Cage
PM1 30653

Technical Definition

Calibrator Set, Radiac, also referenced with federal logistics item name code 00396 - A fixed number of components and/or items, not all having the same basic name, which are required for determination, by comparison, of the radioactivity detecting and measuring accuracy of radiac equipment, such as radiac set. may or may not include comparison of computing accuracy.

Q & A

Part number PM1
What are the historical prices for part number PM1?
No pre-dated pricing for this NSN could be found.
Does this part number PM1 have a shelf life?
No. There is no shelf life applicable for this part.
Does this part number PM1 require Demilitarization?
Non-usml/non-ccli - no demil or dod tsc required. department of commerce may impose licensing requirements to certain destinations. (note 9).
Is part number PM1 ITAR controlled?
No. This item is not ITAR Controlled.
What is the NSN for part number PM1?
6665-01-370-0545
Who are the manufacturers of PM1 and NSN 6665-01-370-0545?
  • Air Force Metrology And Calibration (CAGE 30653)

PM1

Description: Radiac Calibrator Set
Part Number PM1 PM1

Specification

AFHS Accessory Component Quantity
1 and 1 and 1 and 1 and 1 and 1 and 1 and 1 and 1 and 1
AJJW Component Quantity
6
AJKC Supply Items And Quantities
R6130-01-174-0653 power supply 1; 6625-01-251-5909 multimeter 1; digitalthermometer 1; dosimeter 1; electronicbarometer 01; cavity ionization chamber 4; coaxial termination 1; laser measurement system 1; electrometer 1
AKVY Accessory Controlling Agency
Keithley instruments inc; kitrickmanagement co; ruska instrument co; exradin inc; exradin inc; chesapeake laser systems inc
FEAT Special Features
Chambers: a3 3/8 in radius,3.6 vol cc,0.0 r/nc cal factor; a4 - 3/4 in rad,30 cc volume,0.09 r/nc cal; a51-1/8 in radius,100 cc vol,0.03 r/nccal; a6 - 2-1/4 in. rad,800 cc vol; 0.004 r/nc cal; laser: 1-6 meters/sec velocity; 2 g's acceleration; distance .17-30 meters; elevataion porm 60 deg;electrometer: 200 mv-200 v range; 10 uv-10 mv resolution; power supply: 0-3 kv dc output; thermometer: 0.1 deg resolution; displays degrees f. or c.; barometer: 0.5-19 psia,3.5-130 kpaa
FTLD Functional Description
Calibrates gamma radiation sources (standards) model 81-10 on site; electrometer used to measure charge accumulated by ionization chamber; barometer and thermometer measure pressure andtemperature at point of measurement; laser measurement system measures distance from radiation source to detector and used to calibrate radiation track system
National Stock Number (NSN): 6665-01-370-0545