Electrosurgical Apparat Test Set

A washer having a centrally located hole with a solid or laminated cross-section and two flat parallel bearing surfaces. The hole must be 75 percent or less than the largest dimension and the bearing surface finish is greater than 32 microinch (0.8 micrometer). The thickness cannot exceed 25 percent of the largest outside dimension. When the largest outside dimension is less than 1/4 inch or 6mm, the thickness must be 0.006 inches (0.15mm) or greater and for items 1/4 inch (6mm) or larger, the thickness must be 0.016 inches (0.4mm) or greater. The maximum tolerance is applied to dimensions and used as a determining factor. Items with separable lamination use shim. For items fabricated from ceramic or electrically rated materials by specification or industry standards use insulator (1), washer. Excludes:disk, valve, graphite impregnated and lubricated items.

115/230v, 50-60hz, 700 watts; the test set is used to perform automated testing of electrosurgical units for output power, current, crest factor, peak-to-peak voltage, active or dispersive electrode with either opened or closed load circuit, return fault monitor, return electrode monitor, aspen return monitor,and extended load ranges. the unit has a minimum: 24 user-definable autosequences, each capable of 50 individual checks at a desired test load, power level, dial setting, and electrode configuration; 32 user-selectable isolated test loads; manual, user-programmable, and remote (via rs 232c) modes of testing and printing operation; 0.2 volts/amps uncalibrated oscilloscope output; 200 ohms at 12 watts maximum rf leakage test load range, 3 auxiliary modules; patient return contact monitor/fault test module, 25 ohm bipolar expansion module, and 2000 ohm monopolar expansion modular. comes w/hard side carrying case;2 copies operator/service manual, 3 lead test set special features, note: requisitioner must state voltage requirements when ordering. 115 or 230 vac, 50 or 60 hz general characteristics item description, test set electrosurgical apparatus 115/230 v 50/60 hz part name assigned by controlling agency.

NSN 6515-01-438-2409 Datasheet

(NSN) National Stock Number
6515-01-438-2409
(NIIN) National Item Identification Number
01-438-2409
Requires Demilitarization
Yes
(INC) Item Name Code
77777
Date
Thu, Dec 11, 2008
General Characteristics Item Description
note: requisitioner must state voltage requirements when ordering. 115 or 230 vac, 50 or 60 hz
Part Name Assigned By Controlling Agency
test set electrosurgical apparatus 115/230 v 50/60 hz
Special Features
115/230v, 50-60hz, 700 watts; the test set is used to perform automated testing of electrosurgical units for output power, current, crest factor, peak-to-peak voltage, active or dispersive electrode with either opened or closed load circuit, return fault monitor, return electrode monitor, aspen return monitor,and extended load ranges. the unit has a minimum: 24 user-definable autosequences, each capable of 50 individual checks at a desired test load, power level, dial setting, and electrode configuration; 32 user-selectable isolated test loads; manual, user-programmable, and remote (via rs 232c) modes of testing and printing operation; 0.2 volts/amps uncalibrated oscilloscope output; 200 ohms at 12 watts maximum rf leakage test load range, 3 auxiliary modules; patient return contact monitor/fault test module, 25 ohm bipolar expansion module, and 2000 ohm monopolar expansion modular. comes w/hard side carrying case;2 copies operator/service manual, 3 lead test set

Identification

Demilitarization Code
A - Items subject to the export administration regulations (ear) in parts 730-774 of title 15, code of federal regulations (cfr) (ccli or ear99) and determined by dod to present a low risk when released out of dod control. No demil, mut, or end use certificate is required. May require an export licensefrom doc.
Precious Metals Indicator
A - Item does not contain precious metal.
Hazardous Metals Indicator
N - Indicates there is no data in the hmirs and the nsn is in a fsc not generally suspected of containing hazardous materials.
Pseudo Item Name Code
1613593

Cross Reference

CAGE Codes

11332
General Microwave Corporation Div Kratos Microwave Electronics
U.s./canadian manufacturers
4U744
Fluke Electronics Corp
U.s./canadian manufacturers
80009
Tektronix, Inc.
U.s./canadian manufacturers
C1834
Kontron Europe Gmbh
Non-us manufacturer (nato), other foreign organization or codification bureaus

Freight Information

Air Commodity
MZ -
Air Dimension
A - Shipment is not a consolidation and does not exceed 84 inches in any dimension.
National Motor Freight Class Rating
W - Rating variable
Special Handling
9 - Other or no special handling required (sh)
Type of Cargo
Z - No special type of cargo code applicable
Uniform Freight Classification Item Number
00000
Water Commodity
651 - Instruments, medical and surgical
National Motor Freight Classification Item Number
056590

Questions & Answers

  1. What is the FSCG of the NSN 6515-01-438-2409?

    The FSCG of the NSN 6515-01-438-2409 is 6515

  2. Does NSN 6515-01-438-2409 require demilitarization?

    Items subject to the export administration regulations (ear) in parts 730-774 of title 15, code of federal regulations (cfr) (ccli or ear99) and determined by dod to present a low risk when released out of dod control. No demil, mut, or end use certificate is required. May require an export licensefrom doc.

  3. Does NSN 6515-01-438-2409 contain precious materials?

    Item does not contain precious metal.

  4. Does NSN 6515-01-438-2409 contain hazardous materials?

    Indicates there is no data in the hmirs and the nsn is in a fsc not generally suspected of containing hazardous materials.

  5. When the NSN 6515-01-438-2409 was assigned to the Federal Catalog?

    It was assigned 17 years ago, on Thu, Dec 11, 2008

Globe Transparent
Get in Touch
Looking for technical parts, product details, or custom project planning? Our team has you covered with end-to-end support every step of the way.
Contact Us

Pricing & Availability

Send us a request for quote using the form below. Our team will contact you with an up to date pricing and availability quote ASAP.